3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
231
Figure 228.
Removing the Advanced EI Ion Source for Cleaning
12. Twist the Vespel cap off the ion source cartridge and remove the filament. You may now
disassemble the Advanced EI ion source and clean or replace individual parts. See
.
Figure 229.
Advanced EI Ion Source Cartridge
Reassembling the Advanced EI Ion Source Cartridge
The ion source parts are assembled the same way as in
“Reassembling the Extractabrite Ion
up until the filament although you may use gloved hands
instead of the source holder.
1. Add the filament to the correctly assembled Advanced EI ion source cartridge. See
.