Advanced Troubleshooting
Replacing Components of the Analyzer
162
TSQ 9000 Hardware Manual
Thermo Scientific
.
14. Reattach the Q3 exit lens assembly and the Q3 quad assembly to the analyzer tray. Ensure
that quadrupole 3 and the collision cell assembly are correctly positioned before
tightening screws.
15. Reattach the two screws on each side of the quadrupole rod endcaps.
CAUTION - INSTRUMENT DAMAGE
Do not overtighten the endcap screws. If torque
exceeds 1.5 N•m, thread damage will occur.
ATTENTION DOMMAGES À L’INSTRUMENT
Ne serrez pas les vis du bouchon de façon
excessive. Si le couple dépasse 1,5 N•m, le fil sera endommagé
IMPORTANT
For optimal performance, the quadrupole ceramics must fit straight and
securely in the endcap and be aligned so that the collision cell is level with the tray.
Slightly tighten each screw several times going from one to another until all are securely
tightened.
Figure 135.
Reattaching Quadrupole 1, the Ion Guide, and the Collision Cell Assembly to the
Analyzer Tray
Tip
You will get better signal if you gently push the quadrupole rod endcaps toward one
another while tightening the screws.
Tip
Look through the endcaps to be sure the quadrupole ceramics are aligned properly.
Ion Guide
Screws
Q1 Quad
Assembly
Collision Cell