Replacing Components of the Vacuum Interlock
Thermo Scientific
TSQ 9000 Hardware Manual
249
32. Reattach the three screws holding the metal valve body to the front manifold plate.
33. Reattach the two screws connecting the black pillow block to the manifold door.
Figure 261.
Reattaching the Metal Valve Body to the Front Manifold Plate
Figure 262.
Reattaching the Clamp
Screws
Front
Manifold
Plate
Black Pillow Block
Screws
CAUTION INSTRUMENT DAMAGE:
If the vacuum interlock handle is not fully down
and the ball valve is not fully closed, then the ball valve may not be able to completely
open during use and it could damage the ball valve and probes. Confirm the ball is fully
opened and closed as the handle is moved up and down.