Advanced Troubleshooting
Replacing Components of the Source Exchange Tool
Thermo Scientific
TSQ 9000 Hardware Manual
235
To replace the components of the source exchange tool:
1. Use a T20 Torxhead screwdriver to remove the three screws at the end of the black
handle.
2. Loosen the set screw at the other end of the black handle.
3. Slide the black handle toward the other end of the tool.
4. Use a T10 Torxhead screwdriver to remove the set screw holding the endcap onto the
interior shaft.
5. Remove the endcap.
6. Slide the handle off the rest of the tool. Set it aside.
7. Remove the interior shaft from the barrel.
Figure 235.
Removing Source Exchange Tool Screws
Figure 236.
Removing the Endcap
Figure 237.
Removing the Interior Shaft
Screws
Set Screw
Endcap
Set Screw
Interior Shaft
Barrel