Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
317
Figure 338.
Removing the Large Glass Manifold Covers
12. Look into the instrument from the left side and use a T20 Torxhead screwdriver to
remove the three screws around the 4-pin feedthrough.
Figure 339.
Removing the 4-pin Feedthroughs
Large Glass
Manifold Cover
CAUTION - ELECTRICAL HAZARD
Avoid dropping screws between chassis and side cover
or board. If you drop a screw, find it before powering the instrument on again.