![Thermo Scientific TSQ 9000 Скачать руководство пользователя страница 105](http://html.mh-extra.com/html/thermo-scientific/tsq-9000/tsq-9000_hardware-manual_1104856105.webp)
1
Performing Routine Maintenance
Replacing a Dual Filament
Thermo Scientific
TSQ 9000 Hardware Manual
77
14. Remove the filament.
15. Insert the replacement filament into its slot, rotate the spring into position, and slightly
tighten the screw holding the filament retaining clip in place.
16. Bend the filament wires so they do not touch any metal component. You may want the
old filament as a model.
17. Attach the connector of the replacement filament to the source interface board.
Figure 69.
Replacing the Filament
Filament
Filament
Retaining
Spring
Screw
Source
Interface
Board
Содержание TSQ 9000
Страница 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...
Страница 8: ......
Страница 124: ...1 Performing Routine Maintenance Cleaning the Filters 96 TSQ 9000 Hardware Manual Thermo Scientific ...
Страница 476: ......