Advanced Troubleshooting
Replacing Components of the Source Exchange Tool
Thermo Scientific
TSQ 9000 Hardware Manual
237
12. Use the bushing to push the seal back into the barrel until the bushing is back in place.
13. Remove the bushing and replace it, if necessary.
14. Reattach the metal clip into the bottom of the barrel.
15. Insert the interior shaft into the barrel.
16. Slide the handle onto the tool.
17. Attach the endcap and reattach the screw holding it to the interior shaft.
18. Slide the handle onto the endcap.
19. Attach the three screws on the end of the handle.
20. Tighten the set screw.
Figure 240.
Reinserting the Interior Shaft
Figure 241.
Reattaching the Endcap
Figure 242.
Reattaching the Source Exchange Tool Screws
Note
The purpose of the set screw is to keep the components together. Don’t overtighten
it or the handle will be hard to slide up and down the barrel.
Interior Shaft
Endcap
Screws
Set Screw