3
Advanced Troubleshooting
Replacing the Calibration Gas Components
378
TSQ 9000 Hardware Manual
Thermo Scientific
Transfer Line
To replace the transfer line and ferrule
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Open the front door of the GC and disconnect the column and transfer line nut.
3. Move the TSQ 9000 instrument away from the GC so you can access the transfer line on
the right side.
4. Use a T20 Torxhead screwdriver to remove the five screws on the right side panel.
5. Slide the right side panel to the right of the instrument and lift it off.
6. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
IMPORTANT
Be sure to install the transfer line correctly or the ion source will not align
correctly.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
WARNING BURN HAZARD:
The transfer line may be hot, so be careful.
Figure 402.
Removing the Right Side Panel Screws
Note
Remove the screws completely so that they do not scratch the manifold when you
close the door.
Right Side
Panel Screws