Advanced Troubleshooting
Replacing Components of the Ion Source
224
TSQ 9000 Hardware Manual
Thermo Scientific
Figure 219.
Inserting the Small Source Removal Tool
11. Twist the tool to the left. The ion source cartridge should now be connected to the small
source removal tool. See
.
Figure 220.
Twisting the Small Source Removal Tool
12. Slowly remove the ion source cartridge from the instrument.