Performing Routine Maintenance
Replacing a Dual Filament
Thermo Scientific
TSQ 9000 Hardware Manual
75
11. Remove the small glass manifold cover from the top of the instrument.
12. Look through the top of the instrument and use a clean T10 Torxhead screwdriver to
loosen the screw holding the filament retaining clip in place. Then rotate the clip away
from the filament.
Figure 66.
Removing the Small Glass Manifold Cover
IMPORTANT
Many nitrile and latex gloves not certified for clean room use contain
silicone mold releasing agents that will contaminate the instrument. For this reason, clean
room gloves are strongly recommended when touching components inside the vacuum
manifold. We recommend Cardinal Health CP100 Nitrile Cleanroom Gloves. See the
ISQ and TSQ GC-MS Spare Parts Guide
for ordering information.
Small Glass
Manifold Cover