Advanced Troubleshooting
Replacing Components of the Ion Source
210
TSQ 9000 Hardware Manual
Thermo Scientific
23. Let the ion source cartridge cool down before removing it from the source holder.
24. Disassemble the ion source cartridge by removing the locking ring first, then the repeller
spring, then the nut, insulator, and repeller (which comes out in one piece), ion volume,
lens 1, lens 2, and lens 3/RF lens.
25. Set the components on a clean work surface.
26. Replace the ion source cartridge or any individual component.
Reassembling the Extractabrite Ion Source Cartridge
1. Place the ion volume-repeller insulator on the repeller and hold it in place with the
repeller nut. Set it aside for now.
WARNING BURN HAZARD:
The ion source cartridge may be hot, so use caution.
IMPORTANT
Many nitrile and latex gloves not certified for clean room use contain
silicone mold releasing agents that will contaminate the instrument. For this reason,
clean room gloves are strongly recommended when handling the ion source cartridge.
We recommend Cardinal Health CP100 Nitrile Cleanroom Gloves. See the
ISQ and
TSQ GC-MS Spare Parts Guide
for ordering information.
Figure 200.
Extractabrite Ion Source Cartridge Components
Ion Volume
Locking Ring
Repeller
Lens 1
Lens 2
Lens 3/RF Lens
Ion Cartridge Sleeve
Ion Volume
Repeller
Spring
Repeller
Nut
Ion
Volume-Repeller
Insulator