3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
355
Electrometer Board and Shield
To replace the electrometer board and shield
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Disconnect the electrometer board cable.
4. Use a T10 Torxhead screwdriver to remove the two screws holding the electrometer shield
in place.
5. Lift off the electrometer shield and replace it if necessary.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 381.
Disconnecting the Electron Multiplier Cables
Electron
Multiplier Plate
Electrometer
Board Cable
Electrometer
Shield
Screws