1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
58
TSQ 9000 Hardware Manual
Thermo Scientific
5. With the longer teeth of lens 1 facing down toward lens 2, align the larger metal section
of lens 1 with the sleeve window and let it fall into place.
6. Insert the ion volume with the handles fitting into the notches of the sleeve. Make sure
the ion volume is firmly seated into the gap on lens 1. You may need to rotate lens 1
slightly to make the ion volume fit correctly.
Figure 50.
Inserting Lens 1 into the Source Sleeve
Figure 51.
Inserting the Ion Volume into the Source Sleeve
Longer Teeth
of Lens 1
Larger Metal
Section
Sleeve
Window
Ion Volume
Handles
Lens 1 Gap
Handles Fit
Here
Содержание TSQ 9000
Страница 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...
Страница 8: ......
Страница 124: ...1 Performing Routine Maintenance Cleaning the Filters 96 TSQ 9000 Hardware Manual Thermo Scientific ...
Страница 476: ......