Advanced Troubleshooting
Replacing Components of the Manifold
260
TSQ 9000 Hardware Manual
Thermo Scientific
3. Disconnect the vacuum interlock cable from the connector.
4. Pinch together the two springs on the left side the vacuum interlock shield and remove it.
5. Lift the vacuum interlock shield off the instrument. The notch at the bottom of the track
will go around the handle. Be sure to straighten the ball valve handle prior to removing
the VPI shield.
Figure 274.
Disconnecting the Vacuum Interlock Cable
Figure 275.
Removing the Vacuum Interlock Shield
Vacuum
Interlock
Cable
Connector
Springs
Ball Valve
Handle