1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
50
TSQ 9000 Hardware Manual
Thermo Scientific
encounter resistance while removing the source. To release the source from the spring
contacts make small repeated rotations of the handle to the left and right quickly.
12. Once you reach the end of the track (when the back line at the end of the barrel is
uncovered by the black handle), twist the handle to the left to lock it into place.
Figure 38.
Pulling the Tool Down the Left Track
Figure 39.
Locking the Tool in Place
Note: Tool
Rotated Left