Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
185
9. Reattach the thumbscrew-repeller spacers.
10. Reattach the source-repeller thumbscrews to the spacers.
11. Reattach the manifold door and the four screws.
12. Reattach the top cover and tighten the top cover panel screw.
13. Restart your system by following the instructions in
Note
When reattaching the spacers, make sure they fit into the grooves on the
thumbscrews. Also, do not over-tighten the thumbscrews or it may crack the spacers.