Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
250
TSQ 9000 Hardware Manual
Thermo Scientific
34. Reattach the manifold door and the four screws.
35. Hold the spring and safety interlock in place as you attach the vacuum interlock shield
and screws.
36. Reattach the vacuum interlock knob.
Figure 263.
Reattaching the Manifold Door