1
Performing Routine Maintenance
Replacing a Dual Filament
78
TSQ 9000 Hardware Manual
Thermo Scientific
18. Return the small glass manifold cover to the top of the instrument.
19. Swing the manifold door so that it is parallel to the opening. Guide the alignment pins
into the thumbscrews.
Figure 70.
Removing the Small Glass Manifold Cover
Small Glass
Manifold Cover
O-ring
Manifold Door
CAUTION
Be sure the glass manifold cover completely covers the o-ring so that the
manifold is properly sealed. The glass cover must not extend forward over the manifold
door.
ATTENTION
: assurez-vous que le couvercle du collecteur en verre recouvre complètement
le joint torique afin que le collecteur soit correctement scellé. Le couvercle en verre ne doit
pas dépasser la porte du collecteur.
Note
The manifold door will not swing closed. There will be a 1 cm gap. You must push
the door to close it completely. Do not apply excess force to the thumbscrews or
components may be damaged.
Note
Once the manifold door is closed, ensure the filament wires do not contact the
metal surface of the door.