![Thermo Scientific TSQ 9000 Скачать руководство пользователя страница 84](http://html.mh-extra.com/html/thermo-scientific/tsq-9000/tsq-9000_hardware-manual_1104856084.webp)
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
56
TSQ 9000 Hardware Manual
Thermo Scientific
3. Separate the repeller nut, ion volume-repeller insulator, and repeller.
4. Clean the ion source cartridge as instructed in
.
Reassembling the Ion Source Cartridge
1. Place the ion volume-repeller insulator on the repeller and hold it in place with the
repeller nut. Set it aside for now.
2. Insert the ion cartridge sleeve into the source holder.
Figure 47.
Inserting the Sleeve into the Source Holder
Содержание TSQ 9000
Страница 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...
Страница 8: ......
Страница 124: ...1 Performing Routine Maintenance Cleaning the Filters 96 TSQ 9000 Hardware Manual Thermo Scientific ...
Страница 476: ......