3
Advanced Troubleshooting
Replacing Components of the Ion Source
230
TSQ 9000 Hardware Manual
Thermo Scientific
8. Open the manifold door.
9. Let the system cool down for 30 minutes.
10. Disconnect the filament wires from the source interface board.
Figure 227.
Disconnect the Filament Wires
11. Grasp the Vespel cap and gently remove the Advanced EI ion source from the ion source
block. See
.
Figure 226.
Removing the Manifold Door Screws
CAUTION - BURN HAZARD
The ion source cartridge may be hot.
AVERTISSEMENT RISQUE DE BRÛLURE
:. la cartouche de la source d'ions peut être
chaude.
Filament Wires
Содержание TSQ 9000