1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
60
TSQ 9000 Hardware Manual
Thermo Scientific
9. Slide the repeller spring onto the repeller.
10. Place the locking ring on top of the repeller spring so that the repeller protrudes through
the center hole on the locking ring. The hooks on the sleeve fit between the larger gaps on
the locking ring.
Figure 53.
Inserting the Repeller Spring into the Sleeve
Note
Do not twist and lock the locking ring on the sleeve at this time.
Repeller
Spring