Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
329
16. From the left side of the instrument, reattach the three screws around the 20-pin
feedthrough. Tighten the screws uniformly.
17. Look through the top of the instrument and align the pins on the 20-pin feedthrough
with the holes in the source interface board to connect them together.
18. Slide the heat shield into its slot and reattach the screw holding it in place.
19. Reattach the filament board to the source interface board.
20. Close the manifold door and reattach the four screws.
Figure 353.
Reattaching the 20-Pin Feedthrough
20-Pin Feedthrough
Screws
Содержание TSQ 9000
Страница 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...
Страница 8: ......