Replacing Components of the Vacuum Interlock
Thermo Scientific
TSQ 9000 Hardware Manual
251
37. Reconnect the vacuum interlock cable from the connector.
38. Restart your system by following the instructions in
Figure 264.
Reconnecting the Vacuum Interlock Cable
Vacuum
Interlock
Cable
Connector
Содержание TSQ 9000