Advanced Troubleshooting
Replacing the Calibration Gas Components
372
TSQ 9000 Hardware Manual
Thermo Scientific
13. Reach through the left side of the instrument and disconnect the vacuum line tubing that
goes from the foreline connector to the calibration gas module.
14. Pull the tubing out through the access hole in the interior front panel.
15. Disconnect the P1 and P2 electrical wires from the front of the calibration gas controller.
ATTENTION DOMMAGES À L’INSTRUMENT
Limitez au maximum la manipulation des
cartes. Ne touchez aucun composant de carte. Gardez la carte uniquement près du support
en tôle.
Figure 396.
Disconnecting the Calibration Gas Controller
Note
If you are using CI, you will have P1 and P2 electrical wires.
Note
If you have a single-level calibration gas controller, your instrument only has a P1
connector.
P1 and P2
Electrical Wires
Calibration Gas
Module
Foreline Pump Tubing
Access