3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
189
Lens Heater Block Grounding Strap
To replace the lens heater block grounding strap:
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the TSQ 9000 Instrument”
3. Use a T10 Torxhead screwdriver to disconnect the lens heater block grounding strap from
the lens heater block.
4. Attach a new lens heater block grounding strap and reattach the screws.
5. Once the components are correctly reattached, follow the instructions in
Analyzer Tray in the TSQ 9000 Instrument”
to replace the tray in the TSQ
instrument.
6. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 172.
Disconnecting the Lens Heater Block Grounding Strap
Lens
Heater
Block
Lens Heater
Grounding Strap
Screw
Содержание TSQ 9000