3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
338
TSQ 9000 Hardware Manual
Thermo Scientific
5. Locate the PC communication board, which is attached to the controller interface board.
.
6. Disconnect the Ethernet cable from the PC communication board.
7. Use a T20 Torxhead screwdriver to loosen the two M4 screws holding the controller
interface board support bracket to the bottom of the instrument.
ATTENTION DOMMAGES À L’INSTRUMENT
Limitez au maximum la manipulation des
cartes. Ne touchez aucun composant de carte. Gardez la carte uniquement près du support
en tôle.
Note
Although the PC communication board is attached to the controller interface
board, they are sold separately. See the
ISQ and TSQ GC-MS Spare Parts Guide
for
ordering information.
Figure 362.
Locating the Ethernet Cable Connector on the PC Communication Board
Ethernet Cable Connector