Performing Routine Maintenance
Replacing the Electron Multiplier
82
TSQ 9000 Hardware Manual
Thermo Scientific
8. Open the front door of the GC and loosen the transfer line nut. Then pull the column
back (into the oven) about 5 cm to ensure the column is no longer in the ion source.
9. Use a T20 Torxhead screwdriver to loosen the top cover panel.
10. If you are using hydrogen, remove the remaining hydrogen safety screws.
11. Disconnect the electrometer board cable and large feedthrough cables labeled
Dynode
and
Electron Multiplier
.
12. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
ATTENTION DOMMAGES À L’INSTRUMENT
Ne pas poursuivre tant que l’instrument n’a
pas été ventilé, ou des morceaux de la colonne ou de la virole pourraient éclater dans
l’instrument. Pour vous assurer que l’instrument est ventilé, vérifiez dans quelle mesure le
couvercle en verre comprime le joint torique du couvercle supérieur dans le collecteur.
Lorsque la surface du joint torique en contact avec le verre équivaut à environ 1 mm, vous
pouvez sans danger ouvrir l’instrument et enlever la colonne.
Note
You might have to move the GC to access the right hydrogen safety screw.
Figure 73.
Disconnecting the Electron Multiplier Cables
Electron
Multiplier Plate
Dynode
Electrometer
Board Cable
Electron
Multiplier
WARNING - ELECTRICAL SHOCK HAZARD
Avoid dropping screws between chassis and
side cover or board. If you drop a screw, find it before powering the instrument on again.