Upgrade Equipment
Advanced EI Ion Source
420
TSQ 9000 Hardware Manual
Thermo Scientific
Advanced EI Ion Source
If you did not order a TSQ 9000 system with an Advanced EI ion source, you can order a
system later. You must change the source magnets on the manifold door to use an Advanced
EI source. To install an Advanced EI ion source.
1. Follow the steps in
to remove the Extractabrite ion
source cartridge.
2. Follow the steps in
“Powering Off the TSQ 9000 System”
to shut down the
TSQ 9000 system.
3. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
4. Open the manifold door.
Note
Remove the screws completely so that they do not scratch the manifold when you
close the door.
Figure 445.
Opening the Manifold Door
Screws