Advanced Troubleshooting
Replacing Components of the Analyzer
152
TSQ 9000 Hardware Manual
Thermo Scientific
Figure 121.
Replacing the Collision Cell and Quadrupole 3 in the Analyzer Tray
24. Reconnect the three lens wires attached to the lenses between the quadrupoles.
25. Reconnect the Q1 entrance lens wire to the source interface board.
26. Once the components are correctly attached to the analyzer tray, follow the instructions
in
“Replacing the Analyzer Tray in the TSQ 9000 Instrument”
to replace the
tray in the TSQ instrument.
27. Restart your system by following the instructions in
Tip
The endcaps have cutouts that you should use to check to be sure the ceramics are
aligned properly before attaching the two screws.
Collision Cell
Assembly
Q3 Exit Lens
Assembly
Q3 Quad
Assembly
Screws