3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
175
16. If you need to replace the lens heater block, remove the screw holding the lens plate and
springs to the lens heater block.
17. Remove and replace the lens heater block.
18. Attach the lens plate and springs to the lens heater block.
19. Attach the source heater block.
20. Reattach the lens/source heater to the new source heater block and set them next to the
rest of the ion source.
Figure 152.
Replacing the Lens Heater Block
Figure 153.
Reattaching the Lens/Source Heater
Lens Heater Block
Lens Plate
and Spring
Screw
Screws and Cone
Springs
Lens/Source
Heater