Advanced Troubleshooting
Replacing Components of the Analyzer
156
TSQ 9000 Hardware Manual
Thermo Scientific
18.
.
CAUTION - INSTRUMENT DAMAGE
Do not overtighten the endcap screws. If torque
exceeds 1.5 N•m, thread damage will occur.
ATTENTION DOMMAGES À L’INSTRUMENT
Ne serrez pas les vis du bouchon de façon
excessive. Si le couple dépasse 1,5 N•m, le fil sera endommagé
IMPORTANT
For optimal performance, the quadrupole ceramics must fit straight and
securely in the endcap and be aligned so that the collision cell is level with the tray.
Slightly tighten each screw several times going from one to another until all are securely
tightened.
Figure 128.
Reattaching Quadrupole 1, the Ion Guide, and the Collision Cell Assembly to the
Analyzer Tray
Ion Guide
Screws
Q1 Quad
Assembly
Collision Cell