3
Advanced Troubleshooting
Replacing Components of the Ion Source
180
TSQ 9000 Hardware Manual
Thermo Scientific
5. Extract the thumbscrew-repeller spacers and transfer to new thumbscrews.
6. Attach new source-repeller thumbscrews to the ion source.
7. Reattach the manifold door and the four screws.
8. Restart your system by following the instructions in
Figure 160.
Replacing the Source-Repeller Thumbscrews
Source-Repeller
Thumbscrews