Chapter 9. MEMORY CARD OPERATION
9-5 Error Message List
When an error occurs, error messages such as “
Card Error-XX
” (XX denotes error code) are displayed on the
message panel during memory card operations. The table below lists the error codes and explain their meaning.
Memory card operations are aborted when an error occurs.
To return to the normal display mode, press the
DISP key
.
9-12
Code
Meaning
Remedial measures
1
Card insertion failure or card removed
Do over.
2
Card write protect
Replace the card, or reset the protect by SLP550.
3
Card read protect
Replace the card, or reset the protect by SLP550.
4
Bad card
Replace the card.
5
Invalid card format
Initialize the card.
6
Card full
Erase unnecessary files, or initialize the card.
7
Card busy
Do over.
8
File write protect
Initialize the card.
9
Card access error
Do over.
11
Card access sequence error
Do over.
12
FAT abnormal
Initialize the card.
14
Card access sequence error
Do over.
(in file control)
15
Card battery voltage drop (warning) Replace the card, replace the card battery (If
replaceable).
16
Wrong file version
Create new file, and create new data.
17
Data or file are missing.
Create new file, and create new data.
18
DCP551 data full (program load)
Delete unnecessary programs in the DCP551.
19
DCP551 hardware error (load error)
Do over.
20
Card data invalid
Do over.
21
Card data check sum error
Operate the instrument again, or check
(program data)
the program setting of the DCP551.
22 to 36
Card data check sum error
Operate the instrument again, or check
(parameter data)
the parameter setting of the DCP551.
37
Memory protect error
Set DCP551 variable parameter
PA02
to 0 to
(loading the data is protected by the
cancel protection.
variable parameter
PA02
.)
43
No alternate areas remain on the
Replace the card.
EEPROM card
44
Error occurred in writing to alternate Replace the card.
area on EEPROM card
63
Card battery voltage drop (error)
Replace the card, replace the card battery (If
replaceable).
64
File abnormal (card was removed)
Create new file, and create new data.
Содержание DCP550
Страница 1: ...EN1I 6186 Issue 13 04 08 DCP551 Mark ΙΙ Digital Control Programmer User s Manual www honeyvell energy ...
Страница 115: ...Chapter 7 PARAMETER SETUP Settings by event type For information on event operations see Events pages 5 5 to 5 15 7 13 ...
Страница 117: ...Chapter 7 PARAMETER SETUP 0 1 02 3 02 3 0 1 1 4 5 1 4 6 1 4 5 2 2 2 7 15 ...
Страница 119: ...Chapter 7 PARAMETER SETUP 0 11 2 2 3 3 7 17 ...
Страница 122: ...Chapter 7 PARAMETER SETUP d A5 tP A5 CP A5 rE A5 P A6 I A6 d A6 rE A6 CP A6 tP A6 P A7 I A7 d A7 rE A7 CP A7 tP A7 7 20 ...
Страница 209: ...Chapter 12 CALIBRATION Figure 12 11 Current Outputs 12 18 ...
Страница 229: ......
Страница 230: ...No CP UM 5024E ...