Chapter 8 USING OPTIONS AND EXTERNAL DEVICES
8-41
Device Message
No. Function
Header
Numeric
Part
Description
5-14
Fiber end threshold
EFL
(BPL)
m
Sets the fiber end threshold.
3
≤
m
≤
10:1step
5-15
Fault event show/hide
setting
DFE
m
Shows/hides fault events.
m
=
0: Shows fault events.
1: Hides fault events.
5-16
Fault event threshold
(Splice loss)
FESL
m
Sets the splice loss threshold for fault events.
0.01
≤
m
≤
9.99:0.01step
5-17
Fault event threshold
(Return loss)
FERL
m
Sets the return loss threshold for fault events.
20
≤
m
≤
70:1step
5-18
Auto search event count
output
No control command
5-19
Auto search result output
No control command
5-20
Section analysis start point
setting
SSPOS
Sets the start point at the cursor position.
5-21
Section analysis end point
setting
SEPOS
Sets the end point at the cursor position.
5-22
Reference return loss
setting
AJPOS
Sets the cursor position as the reference return
loss.
5-23
Section data acquisition
No control command
6-1
Drive setting
FDA
m, n
Sets the drive and file No.
m
=
0: FD (expansion unit)
1: Built-in memory
2:
-
3:
PCMCIA
4:
-
5:
USB
n
=
File name (excluding file extension)
6-2
Folder setting
DIR
m
Sets the folder name.
The device message must be separated from the
originating location using a comma “,”
m
=
Folder name (root folder sets “/”)
example:/aaa/bbb/ccc/ddd/
6-3
File type setting
FF
m
Sets the file type.
m
=
0: .TRB
6: .LST
1: -
7: .SOR(Telcordia)
2: .SET
8: .CSV
3: .SOR(Bellcore) 9: .TRD
4: .TIF
10: .TRA
5: .BMP
11: .4
(In file saveing, 3 and 7 will be not made
distinction)
6-4
Recorded capacity
No control command
6-5
Recorded file name
No control command
6-6
Sub folder check
No control command
6-7
Acquisition of file list of
current folder
No control command
7-1
File recall
FRC
Recalls a file.
.BMP, .TIF, .LST, and .CSV files cannot be
recalled.
Drive (6-1), folder (6-2) and file type (6-3) must be
specified in advance.
Summary of Contents for AQ7260 OTDR
Page 1: ...User s Manual Y okogawa Electric Corporation AQ7260 OTDR IM 813920300 01E 2nd Edition ...
Page 27: ......
Page 70: ...Chapter 2 BEFORE STARTING MEASUREMENT 2 1 Changing System Settings 2 2 ...
Page 76: ...Chapter 2 CHANGING SYSTEM SETTINGS 2 7 When LINE is selected When DOT is selected TIP ...
Page 78: ...Chapter 2 CHANGING SYSTEM SETTINGS 2 9 When CROSS is selected When LINE is selected TIP ...
Page 89: ...Chapter 2 CHANGING SYSTEM SETTINGS 2 20 When LINE is selected When ARROW is selected TIP ...
Page 216: ...Chapter 4 ENTERING CHARACTERS 4 1 Entering Characters 4 2 4 2 Editing Characters 4 7 ...
Page 264: ...Chapter 6 FILE OPERATION 6 1 File Operation 6 2 6 2 Using the Utility Functions 6 22 ...
Page 295: ......
Page 351: ......
Page 413: ......
Page 422: ...Chapter 9 SPECIFICATIONS 9 9 AQ7260 OTDR ...
Page 423: ...Chapter 9 SPECIFICATIONS 9 10 AQ7261 SMF MODULE ...
Page 424: ...Chapter 9 SPECIFICATIONS 9 11 AQ7264 SMF MODULE ...
Page 425: ...Chapter 9 SPECIFICATIONS 9 12 AQ7265 SMF MODULE ...
Page 426: ...Chapter 9 SPECIFICATIONS 9 13 PRINTER FDD UNIT ...
Page 427: ...Chapter 9 SPECIFICATIONS 9 14 PRINTER UNIT ...
Page 428: ...Chapter 10 APPENDIX 10 1 Software Upgrade 10 2 10 2 Troubleshooting 10 5 10 3 Glossary 10 8 ...