Installing Optional Kits
Installing the NoVent Microfluidics
376
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing the NoVent Microfluidics
This section provides instructions for installing the NoVent™ Microfluidics on your TRACE
1300 or TRACE 1310 GC, and your ISQ or TSQ 8000 Series mass spectrometer.
Connecting the NoVent Microfludics Module to the TRACE 1300/1310
According to the configuration of your TRACE 1300 or TRACE 1310 GC, a dummy
module is present in the free site where the detector module is not installed.
If no dummy module is present, and no vacant module position is present on either the
TRACE 1300 or TRACE 1310 GC or on the optional gas valve oven (if you have one), you
must take one of the existing and installed modules off in order to install the NoVent
Microfluidics.
Figure 372.
Location of Detector Modules on your TRACE 1300/TRACE 1310 GC
Contents
•
“Connecting the NoVent Microfludics Module to the TRACE 1300/1310”
•
“Installing the Mounting Bracket”
•
“Preparing the NoVent Microfluidics Restrictor Tubing”
•
“Attaching the Ferrule and Nut to the GC Column”
•
“Attaching the New Tubing to the Transfer Line”
•
“Connecting the Capillaries to the Microfluidics Splitter”
•
•
Dummy Back Detector Module
Dummy Front Detector Module