Detectors Advanced Maintenance
Cleaning or Replacing the FID Jet
272
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 249.
FID Cables
Figure 250.
FID Cables Disconnection
a.
Loosen the glow-plug cable socket set screw using a T10 Torxhead screwdriver.
Carefully pull out the terminal body of the glow-plug cable from the dowel pin on
the glow-plug element.
b. Unscrew and pull out the coaxial straight plug crimp connector of the signal cable
from the collecting electrode bulkhead jack.
c.
Pull out the quick coupling coaxial straight jack connector of the polarizing cable
from the polarizing electrode bulkhead jack.
d. Carefully move the cables in order to have free space for handling the detector.
8. Remove the top parts of the detector. See
.
Glow-Plug Cable Terminal body
Polarizing Cable Straight Jack Connector
Signal Cable Straight Plug Crimp
Connector
Socket Set Screw
Polarizing Electrode Bulkhead Jack
Collecting Electrode Bulkhead Jack
Glow-plug Dowel Pin