12
Troubleshooting
Error Messages
584
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Error Messages
Error messages are visualized in case of GC malfunctioning. See the following sections:
•
•
TRACE 1300 Error Messages
TRACE 1300
instrument malfunction, due to a component failure or to abnormal operating
condition is identified by the blinking of both
Power
and
Ready
lights located on the
TRACE 1300 Status Panel. See
Figure 642.
TRACE 1300: Alarm Identification
A typical sound (beep) is heard. Power and temperatures are automatically cut off.
To reset an alarm, the GC must be powered off, and then powered on.
TRACE 1310 Error Messages
TRACE 1310
instrument malfunctioning, due to a component failure or to abnormal
operating condition, is identified by an error messages displayed before or during the runs.
• FID Front/Back/Left/Right
–
Unconnected
–
Temperature over limit
–
Opened PT100
–
Shorted PT100
–
Thermal safety: Not Heating or Auto Heating
–
Reset detected
• ECD Front/Back/Left/Right
–
Unconnected
–
PLD error
–
Temperature over limit
IMPORTANT
When an alarm is displayed in the message bar, try to solve the
problem, then go to the
Diagnostics
menu and press the
Reset
button to reboot
the system. If the problem persists, contact the Technical Support; see
.