9
Adding Modules
Adding a Generic Detector Interface
476
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Replacing the Encapsulated Flow Restrictors
This section provides instruction for replacing the 1/4-in. OD x 1/4-in. length encapsulated
flow restrictors. The flow restrictors are located inside the manifold marked
Gas Outputs
on
the GDI electrical interface. See
Figure 501.
Gas Outputs Manifold
To replace an encapsulated flow restrictor
1. Remove the Gas Outputs manifold from the GDI electrical interface by unscrewing its
fixing screws using a T20 Torxhead screwdriver. See
Figure 502.
Gas Outputs Manifold
2. Remove the encapsulated flow restrictor of interest from its seat using tweezers.
See
Figure 503.
Flow Restrictor Replacement
Gas Outputs
Manifold
Manifold O-Rings
Pay attention to not lose
the o-ring placed into the
manifold
Internal O-Rings
Pay attention to not
lose the internal
o-ring
Encapsulated
Flow
Restrictors