10
Adding Systems
Adding an Auxiliary Gas System
528
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 573.
Perform the Duct for the Auxiliary Gas Interface (4)
4. Jump to the section
“Installing and Connecting the Auxiliary Gas Module”
.
Installing the Auxiliary Gas Interface on the Right Wall of the Oven
To install the auxiliary gas interface on the right wall of the oven
1. Make sure that the preliminary operations have been carried out. See
2. Remove the right side panel.
a.
Open the front door of the GC.
b. Using a T20 Torxhead screwdriver, unscrew the left side panel screw from the interior
front panel. See
. Save the screw because it will be reused later.