Performing Detectors Routine Maintenance
Maintaining a Flame Ionization Detector (FID)
140
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
The FID detector periodic maintenance includes:
• Cleaning the collecting electrode
It is suggest to clean the collecting electrode annually, or when a high noise baseline is
found, due to some compounds that produce detector contamination. If necessary,
replace the collecting electrode.
See
“Cleaning or Replacing the FID Collecting Electrode”
• Replacing the FID ignition glow-plug
This operation is NOT part on the ordinary maintenance. The flame ignition element
must be replaced only when defective.
See
“Replacing the FID Ignition Glow-plug”
Before maintaining the detector, read the following precautions:
WARNING
Carry out all the operations at low temperature to avoid burns. Therefore,
before beginning maintenance, cool the detector to room temperature.
CAUTION
When handling organic solvents you must take precautions to avoid health
hazards.
Materials needed to maintain a FID detector
Ultrasonic cleaner
Liquid detergent
GC-grade methanol
Distilled water
FID Ignition Glow-plug
Collecting electrode (if necessary)
8-mm elbowed box wrench
T6 Torxhead screwdriver
Forceps or tweezers
Note
For maintaining or replacing any other component not listed in this section, see
Chapter 7, “Detectors Advanced Maintenance.”
•
“Removing/Replacing a Detector Module”
•