9
Adding Modules
Adding a Generic Detector Interface
472
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Adding a Generic Detector Interface
This section provides instructions for installing the detector and control modules of the
Generic Detector Interface on your TRACE 1300/1310 GC or TRACE 1310 Auxiliary
Oven.
Preliminary Operations
Before starting, the following preliminary operation must be carried out.
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
3. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
4. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
Getting Started
For properly installing the Generic Detector Interface see the following sections:
•
•
“Assembling the GDI Electrical Interface”
•
“Replacing the Encapsulated Flow Restrictors”
•
“Installing and Connecting the GDI Electrical Interface”
•
“Connecting a GDI Interface to the TRACE 1310 Auxiliary Oven”
WARNING
This operation must be carried out by authorized and trained Thermo Fisher
Scientific Service Field Engineers.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
CAUTION
If the GC is coupled to a TRACE 1310 Auxiliary Oven, unplug the power
cable from the AC Input connector into the back of the TRACE 1310 Auxiliary Oven
and from the wall outlet.