8
Installing Optional Kits
Performing the Dual FPD Detector Configuration
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
411
a.
Disconnect the signal, excitation voltage, and ignition/heating cables from the
detector. See
Figure 421.
FPD Cable Disconnection
b. Using the tool provided with the system, loosen the fixing nut on the base of the
.
Figure 422.
FPD Detector Removal
Note
Do not lose the aluminium ring inserted between the detector head and
the base body.
Signal Cable
Excitation
Voltage
Cable
Ignition/Heating Cable
FPD Fixing Nut
Aluminium Ring