10
Adding Systems
Adding the Oven Cryo System
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
505
c.
Fix the solenoid valve assembly to the back of the GC using the two fixing screws
previously removed. See
Figure 543.
Solenoid Valve Assemble Fixing
10. Connect the cryogenic tank tube to the solenoid valve assembly.
a.
Connect the proper end of the cryo supply tube to the 1/8-in. NPT connection of
the solenoid valve using the proper nut and ferrule. Use a 7/16-in. wrench for
tightening the fittings. See
Solenoid Valve
Assembly
Fixing Screws