9
Adding Modules
Adding a PDD Module
468
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
2. Connect the pulses cable.
a.
Plug the 4-pin pulses cable connect into the 4-pin connector on the pulse generator
See
Figure 492.
Pulses Cable Connection (1)
b. On the pulse generator control side, loosen the right nut that fix the pulse generator
to the bracket. Do not loosen the left nut, keep it tightened. Place the safety cover
until the nuts couple the slots on the bottom of the safety cover payning attention to
guide the cable into the slot located on the bottom of the safety cover. See
Figure 493.
Pulses Cable Connection (2)
c.
Tighten the right nut using a T30 torxhead screwdriver. See
.
Slot