6
Injectors Advanced Maintenance
Cleaning the HeS-S/SL Injector Body
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
243
Figure 223.
Instant Connect Helium Saver Injector Module: Cleaning the Injector Body (1)
9. Disconnect the helium transfer tube See
a.
Loosen the captive screw of the helium transfer tube and remove the tube from the
gas delivery block.
Figure 224.
Instant Connect Helium Saver Injector Module: Cleaning the Injector Body (2)
10. Remove the body head o-rings. See
.
a.
Using tweezers, remove both the internal and external body head o-rings.
Septum Holder/Liner Cap with Septum
Liner Seal
Ring Nut
Liner
Septum Cap
Helium Transfer Tube
Gas Delivery Block