Installation
Coupling to a Mass Spectrometer
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
31
Figure 40.
Transfer Line MS Heaters
7. Tune and set the DFS, IRMS, and ICP-MS working conditions of the high resolution
mass spectrometer, referring to the
relevant manuals.
Setting Handshake Parameters
When your GC is connected to a mass spectrometer, set the handshaking parameters as shown
in
To configure handshaking parameters through the touch screen
1. In the main menu select the
Configuration
icon, the Configuration menu appears.
2. In the Configuration menu, select the
Handshake
icon to open the relevant submenu.
3. Set the
Handshaking parameters
as required, then return to main menu.
To configure handshaking parameters through the Chromatography Data System
1. Launch the Data System. In the relevant Configuration page specify the handshaking
parameters.
Heater Connection for
Transfer Line 1
Heater Connection for
Transfer Line 2
Table 1.
GC Handshaking Parameters When A Mass Spectrometer is Connected
Parameter
Set To:
Remote Start In
High to Low
Inhibit Ready In
When High
End of Run Out
High to Low
Start of Run Out
High to Low
GC Ready Out
When Low
Prep Run Out
When Low