Detectors Advanced Maintenance
Cleaning or Replacing the FPD Filter-side Heat Shields
304
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
a.
Insert a new filter-side heat shield into its housing inside the filter support.
See
Figure 292.
Filter-side Heat Shield Replacement (1)
b. Insert the filter support into the flange and fix it tightening the three Allen screws.
Check the o-ring that fixes the heat shield is in place before tightening the screws.
See
Figure 293.
Filter-side Heat Shield Replacement (2)
10. Remount the filter and the photomultiplier assembly.
a.
Reinsert the filter into the support. The mirror face must be oriented towards the
flame. See
.
Figure 294.
Filter Remounting