1
Installation
Making the Gas Supply Plumbing Connections
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
9
b. If present, remove the cap from the gas inlet port to connect. See
.
Figure 6.
Cap Removal
c.
If not already installed, screw the gas inlet fitting provided into the gas inlet port
interposing the o-ring. See
Figure 7.
Gas Inlet Fitting Installation
d. Connect the gas line to the front/back carrier gas inlet. Use a 7/16-in. wrench for
tightening the fittings. See
.
Figure 8.
Carrier Gas Line Connection
e.
Connect the gas line to the front/back detector gas inlet. Use a 7/16-in. wrench for
tightening the fittings. See
,
,
, and
to properly
connect each detector.
CAUTION
Inside each 1/8-in. inlet manifold is a fritted filter. To keep the gas line from
touching, and possibly damaging the filter, extend the tubing only 5-mm past the front
ferrule. This ensures the tubing does not touch the filter.
Gas Inlet Port
Cap
Gas Inlet Fitting
O-Ring
Front Carrier Gas Inlet
Back Carrier Gas Inlet