2
Performing Routine Maintenance
Read Me First
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
67
IMPORTANT
All the maintenance operations must be carried out at low temperature to avoid burns.
Therefore, before beginning the sequence, the GC oven, injectors and detectors must be
cooled to room temperature, and then the gases supply must be closed.
You can carry out these operations manually, or by pressing the
Maintenance
key. For
details refer to the
TRACE 1300 and TRACE 1310 User Guide
.
At the end of the maintenance operations, restore the GC normal working conditions.
Maintenance of an injector
Before opening the injector for maintenance, turn the carrier gas off, and wait for the
carrier pressure to go to zero.
When an autosampler is present:
Move the autosampler away from the injector module to create free space around it:
• If an AI 1310/AS 1310 - AI 3000/AS 3000 II is installed, pull the sampler support
plate outwards.
• If a TriPlus RSH, TriPlus 100 Liquid Sampler, or TriPlus is installed, make sure that
the standby turret position does not obstacle the injector or detector module to
maintain. If yes, change the position of the turret/head.