6
Injectors Advanced Maintenance
Cleaning the PTV Injector Head Assembly
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
247
b. Use tweezers to remove the liner with the liner seal (o-ring) from the injector.
Figure 226.
PTV Top Parts and Liner Removal
9. Remove the fan assembly.
The fan is fixed on a support. Do not remove the fan from its support but remove the
assembly proceeding as follows:
a.
Using tweezers and a T10 Torxhead screwdriver, unscrew the left fixing screw of the
support, then unscrew the right fixing screw. See
.
CAUTION
Be careful not to break the glass liner when removing it. Glass
splinters might fall into the lower part of the vaporization chamber. If the glass
liner breaks, follow the instructions for
“Replacing the PTV Broken Liner”
.
Septum Cap
Septum
Liner Cap
Liner Seal (O-ring)
Liner
Injector Head Assembly